FM200
Laboratory fume cupboard monitoring according to DIN EN 14175
Product Description
The FM200 fume cupboard monitoring system is used as a monitoring and alarm system for exhaust air volume in various applications, such as fume cupboards, safety cabinets and other extraction units. It fulfills the DIN 12924 and DIN EN 14175-2 standards. This means safety for the laboratory technician.
Laboratory fume cupboard monitoring is suitable for all laboratory fume cupboard designs, which makes it easy to implement new installations and retrofits on existing fume cupboards.
Standard version:
- FM200-K Constant monitoring of the air volume, incl. differential pressure sensor
Other versions available:
- FM200-F Constant monitoring of face velocity, incl. air flow sensor, without differential pressure sensor
- FM200-W Constant monitoring air volume, incl. sash position sensor and differential pressure sensor for additional signaling opening height > 50cm on the function display
FM200-FW Constant monitoring by calculating the air velocity, incl. sash position sensor and differential pressure sensor
Special Features
Microcontroller Laboratory Fume cupboard Monitoring System |
Low cost system suitable for mounting on electric DIN rail |
Power supply 24VDC or via separate power plug |
Parameterization and retrieval of all system values via PC4500 software |
Monitoring of supply air and exhaust air systems |
Differential pressure sensor with long-term stability |
Monitoring of the fume cupboard operation according to DIN EN 14175 with acoustic and visual alarms |
Optional monitoring for exceeding a parameterizable volume flow Vmax with visual warning message |
Visual and optionally acoustic warning message for the operating status “Sash opened > 50cm open” |
Parameterization of a second monitoring value Vmin (reduced air volume for night operation) |
Suitable for all laboratory fume cupboard types |
Functional Description
The microcontroller safety system is used to monitor the operating status of fume cupboards. An acoustic and visual alarm is activated as soon as the exhaust air volume falls below or exceeds the parameterizable limit values. For precise and safe monitoring, a suitable measuring system is necessary. For safe operation with the internal differential pressure sensor, as well as reproducible and accurate measurement results, we therefore recommend the venturi measuring device or the measuring tube from SCHNEIDER.
Function display
Each fume cupboard monitoring system includes a function display that is installed in the side panel of the fume cupboard. This is used to show the correct operation of the fume cupboard and the interaction with the user.
An overview of the available function displays can be found in the document “Technical Data Sheet Function Displays FA-100ABCDEFGHIJK”.